Description
CONTROLS have long tradition in manufacturing automatic setting time testers for cement and mortar. A new generation, the fourth one, is now available, featuring a large color 5.1” touch screen display. As with the older generations the test procedure is unchanged, a needle (or a probe) drops freely into a cement sample at regular intervals and in fixed positions. Penetration depth is measured by a sensor with 0,1 mm resolution. Along with hardening process development the penetration depth decreases, when it matches some thresholds pre-defined by Standards initial and final setting times are measured and recorded.
Technical Specifications
Technicalspecifications
- Conforming to EN 196-3, 13279-2, 480-2, ASTM C191, C187
- 5.1”, 800 x 480 pixel, 16 M colors, icon-driven touchscreen graphic display, showing data and plots
- LAN port for direct connection to PC of a single unit or connection to a LAN hub for creating a network with up to 32 independent units all controlled by a single PC. 1 LAN cable is included
- USB port for test data download to PC and easy firmware upgrade
- Unlimited storage capacity for test data on internal 16 GB SD card
- Test procedures can be customized and stored to match user-defined requirements
- Can incorporate an integrated graphic printer showing test result and setting time plot
- Large test space with easy accessibility
- Automatic calculation of initial and final setting time at programmable penetration depth limits
- Wide range of accessories including EN and ASTM/AASHTO parts, in-water testing kit, needle cleaning device, integrated printer, probes for testing consistency and gypsum
- Minimum penetrations rate: 10 seconds
- Penetration measurement by encoder
- Power: 50 W
- Dimensions: 200x400x410mm (L x W x H)
- Weight approx.: 10 kg
Firmwarespecifications
- Easy programming of customized test profiles, recallable for future tests, including:
- adjustable test start delay
- penetration points positions
- manual or automatic penetration rate
- free or driven dropping mode
- holding intervals inside the sample
- automatic end- test detection
- automatic measurement of initial and final setting time
- Test data: test number, operator, client, date, hour, cement type, water percentage, delay
- Easy calibration menu
- Clock calendar
- Multi-language
Download
相关标签
Controls, Vicat Apparatus